CPA100 Capacitive Displacement system
The Lion Precision CPA100 Capacitive Displacement Sensor provides repeatable noncontact measurement of position changes of conductive targets; the CPA100 can also detect nonconductive targets in many applications.
The CPA100 provides a nonlinear analog voltage proportional to changes in the target position and a digital switched (setpoint) output.
Multiple Sensors
When multiple sensors are used with the same target, the sensors must be synchronized for best performance. A synchronization kit includes a cable and “T” connector for each channel. If sensors are not synchronized, interference will increase output noise, especially at lower frequencies.
Maximizing Performance - Ungrounded Targets
Ungrounded targets have the potential to inject noise into the sensor. If the output is unusually noisy, be sure the target is grounded. On moving/rotating targets this can be accomplished with a small metal brush or thin piece of metal which is connected to ground.

1) Actual values depend on probe and range which are listed on the calibration certificate shipped with the product
CPA100 probe range & resolution